Cables and Wiring
uLan communicating modules/instruments are interconnected through two conductor twisted or coaxial cables. The BINDER subminiature circular connectors series 719 are used for connection to the instruments. PC RS-485 communication boards use their specific connectors.
![]() |
Binder 719 series connectors Type 09 9765 2004 male socket is used on instruments Type 09 9764 7004 female insert is used on cables
|
![]() |
Tedia PCI-1482 RS-485 card Card has to be switched for RS-422 Multimaster mode (SW1 1-OFF 2-ON)
|
![]() |
PiKRON ISA RS-485 card The ISA card needs to be switched for correct I/O address and interrupt source by jumpers. The jumpers description in a provided diagram
|
Auxiliary I/O connector
The most of LC5000 instruments is equipped with auxiliary I/O interface for sensing and control of TTL (+5V based) logic signals. The interface can be used for starting of gradient program and data acquisition by a microswitch on injection valve and for interfacing instruments with instruments without own intelligence and communication infrastructure (simple fractions collectors, technological valves, etc.). It can be used for notification of start for third party integration software as well.
![]() |
|